Agilent 8563EC spectrum analyzer |Agilent 8563EC
Frequency range (internal mixing):
Agilent HP8560EC: 30 Hz connection scans to 2.9, 6.5, 13.2, 26.5, 40 or 50 GHz
Agilent HP8561EC: 30Hz~6.5GHz
Agilent HP8562EC: 30Hz~13.2GHz
Agilent HP8563EC: 9kHz~26.5GHz; 30Hz~26.5GHz (option 006)
Agilent HP8564EC: 9kHz~40GHz; 30Hz~40GHz (option 006)
Agilent HP8565EC: 9kHz~50GHz; 30Hz~50GHz (option 006)
频率范围(外混频):
18GHz~325GHz is divided into 12 waveguide frequency bands (cannot be provided by 8560EC option 002)
General Specifications of Agilent HP8563EC Spectrum Analyzer
Environmental Specifications of Agilent HP8563EC Spectrum Analyzer
Calibration cycle: 2 years (8560EC/61EC/62EC/63EC); 1 year (8564EC/65EC)
Warm-up time: 5 minutes under normal environmental conditions
Temperature: -10°~+55℃ (when working); -51°~+71℃ (when not working)
Humidity: 95%, 5 days at 40℃
Water resistance: impervious to 16 liters/hour/ft2 of rain
Height: 15,000 feet (when working); 50,000 feet (when not working)
Impulse shock (half sine): 30g, lasting 11ms
Transport trailer: 8-inch trailer with 6 sides and 8 corners
Electromagnetic compatibility: Conducted and radiated interference comply with CISPR Pub.11 (1990). Except for some indicators, it meets the MIL-STD-461 C part4 standard.
Power requirements
Working with 115Vac: 90~140V rms, 3.2A rms max, 47~440Hz
Work with 230Vac: 180~250V rms, 1.8Arms max, 47~66Hz
Power consumption; 180W (8560EC/61EC/62EC/63EC); 260W (8564EC/65EC)
Acoustic noise (rated value): at room temperature, intensity <5.0 Bel (ISODP 7779)
Dimensions (without handle, bracket, mask): 337mm (width) × 187mm (height) × 461mm (length)
Weight (handling weight, rated value): 20kg
Option 002 built-in tracking generator (only for 8560EC)
Frequency range: 300KHz~2.9GHz
Accuracy (after taking the peak value): ±(reference frequency accuracy×tuning frequency+5%×interval+295Hz)
Tracking drift (rated value): After 5 minutes of warm-up, it can be used for 1KHz RBW: After 30 minutes of heat prevention, it can be used for 300Hz RBW
Small RBW: 300Hz amplitude
Output level: -10~+1dBm; -10~2.8dBm (typical value) Resolution: 0.1dB accuracy (25℃±10℃)
Fine adjustment: ±0.2dB/dB, ±0.5dBmax
Accuracy: 0.75dB
Level flatness: ±2.0dB
Dynamic range: 95dB(300KHz~1MHz); 115dB(1MHz~2.0GHz); 110dB(2.0GHz~2.9GHz)
Power scan: 10dB range, 0.1dB resolution input/output
RF output (front panel): N type (female), 50Ω (rated)
External ALC input (rear panel): BNC (female), used with negative detector